Unexpected high failure rate of a specific MicroPort/LivaNova/Sorin pacing lead
Andreas Haeberlin1, Marie-Theres Anwander2, Thomas Kueffer2
1Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland; Sitem Center for Translational Medicine and Biomedical Entrepreneurship, University of Bern, Bern, Switzerland; Bordeaux University Hospital (CHU), Cardio-Thoracic Unit, Pessac, France.
Background:
Pacing leads are the Achilles heel of pacemakers. Most manufacturers report a 3-year survival rate of >99% of their leads. We observed several failures of the Beflex/Vega leads (MicroPort, Shanghai, China; formerly Sorin/LivaNova).
Objective:
The purpose of this study was to investigate failure rates of Beflex/Vega leads.
Methods:
We analyzed the performance of Beflex/Vega leads implanted at our tertiary referral center. All-cause lead failures (any issues requiring reinterventions such as lead dislocations, cardiac perforations, and electrical abnormalities) were identified during follow-up. The Beflex/Vega lead was compared with a reference lead (CapSureFix Novus 5076, Medtronic, Minneapolis, MN) implanted within the same period and by the same operators.
Results:
A total of 585 leads were analyzed (382 Beflex/Vega and 203 CapSureFix Novus 5076 leads). Cumulative failure rate estimates were 5.2%, 6.3%, and 12.4% after 1, 2, and 3 years for the Beflex/Vega lead. This was worse compared to the reference lead (1.5%, 1.5%, 3.7% after 1, 2, and 3 years; P = .001). Early failure manifestations up to 3 months occurred at a similar rate (Beflex/Vega vs CapSureFix Novus 5076 lead: 1.3% vs 0.5% for dislocations; 1.3% vs 1.0% for perforations). During follow-up, electrical abnormalities such as noise oversensing (P = .013) and increased pacing thresholds (P = .003) became more frequent in the Beflex/Vega group. Electrical abnormalities were the most common failure manifestation 3 years after implantation in this group (9.4% vs 2.2% for the CapSureFix Novus 5076).
Conclusion:
The failure rate of the Beflex/Vega lead of >10% after 3 years was higher than that of a competitor lead. This gives rise to concern since >135,000 such leads are active worldwide.
More Related Videos
10:05Testing the Efficacy of Pharmacological Agents in a Pericardial Target Delivery Model in the Swine
Published on: July 7, 2016
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Cardiac Catheterization IV: Nursing Management
Heart Failure Drugs: Inotropic Agents
Heart Failure II: Pathophysiology
Errors occurring during blood pressure monitoring
Several factors...
Cardiac Catheterization III: Left Heart Catheterization
Heart Failure IV: Classification and Diagnostic Evaluation