Digital Subtraction Myelography is Associated with Less Radiation Dose than CT-based Techniques
Patrick J Nicholson1, William C Guest2, Monique van Prooijen3
1Department of Medical Imaging, Division of Neuroradiology, Toronto Western Hospital, University Health Network, 399 Bathurst, M5T 2S8, Toronto, ON, Canada. patrick.nicholson@uhn.ca.
Purpose:
Both CT myelogram (CTM) and digital-subtraction myelogram (DSM) can be used to evaluate patients for possible cerebrospinal fluid (CSF) leaks. DSM is a relatively new technique. No data exists on the radiation dose associated with this procedure, and how it compares with CTM.
Materials And Methods:
All patients who underwent DSM for spontaneous intracranial hypotension (SIH) refractory to blood patching from Dec 2016 - Sept 2019 were retrospectively assessed. DSM dose factors were then recorded (cumulative fluoroscopy time, total kerma area product (KAP, mGy.cm2), cumulative air kerma (mGy), as well as CTM dose factors (included CTDIvol (mGy) and dose-length product (DLP, mGy.cm). These indices were then used to calculate the effective dose for both procedures using standardized conversion factors.
Results:
61 DSMs were performed in 42 patients, 33 of which also underwent CTM. The median effective dose was 6.6 mSv per DSM study (range: 1.2 - 17.7). On a per-patient basis (i.e. those patients who underwent more than one DSM (as the initial one was negative), the median total effective dose was 13 mSv for their total DSM imaging (range: 2.6 -31.7). For the CTM, the median effective dose was 19.7 mSv (range: 3.2 - 82.4 mSv).
Conclusion:
The radiation dose with DSM appears to be significantly lower than that of CTM (p = 0.0005), when looking at CTM doses both from our institution and in the published literature.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Imaging Studies for Cardiovascular System V: CT
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT