Related Experiment Video
Updated: Dec 11, 2025

Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
A Newly Developed Chest Drainage Unit with an Integrated CO2 Detector
Jens Eckardt1, Marike Lijkendijk1, Peter B Licht1
1Department of Cardiothoracic Surgery, Odense University Hospital, Odense, Denmark.
Purpose:
Postoperative air leakage does not always originate from parenchymal defects. In some cases, it may arise from defects in the chest drainage unit itself or connections, or from reverse airflow in water seals. The aim of the present study was to test a new chest drainage unit in the clinic, where an integrated CO2-sensitive colour indicator helps to distinguish false air leakage from true air leakage.
Methods:
Over a 3-week period, 14 consecutive patients were operated upon using either an open approach (for bilobectomies) or VATS (for diagnostic procedures or lobectomies). All patients received general anaesthesia with double-lumen intubation. All patients had a 24-Fr chest tube connected to a chest drainage unit with a built-in CO2-detector.
Results:
In all patients with air leakage after surgery, we found a colour change in the CO2-sensitive colour indicator, confirming "true air leakage". One patient had prolonged air leakage. None of the patients had pneumothorax after removal of the chest tube and no patients had wound infections or any other complications.
Conclusion:
The chest drainage unit described here was easily implemented in the clinic and clearly confirmed true air leakage in all patients with air leakage after lung surgery. It allowed safe and appropriate timing for chest tube removal with no need for reinsertion in a broad cohort of patients referred for thoracic surgery.
More Related Videos
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Pneumothorax-II
Clinical Manifestations: