Effect of Elevated Ambient Temperature on Simulator-Derived Oscillometric Blood Pressure Measurement
Jennifer S Ringrose1,2, Michael D Kennedy3, Jalisa Kassam4
1Department of Medicine, University of Alberta, Edmonton, Alberta, Canada.
Background:
Oscillometric blood pressure (BP) devices are typically labeled for use up to 40 °C. Many geographic regions have ambient temperatures exceeding 40 °C. We assessed the effect of increased ambient temperature (40-55 °C) on simulator-derived oscillometric BP measurement.
Methods:
Three Omron BP769CAN devices, 3 A&D Medical UA-651BLE devices, and accompanying cuffs were used. A custom heat chamber heated each device to the specified temperature. A noninvasive BP simulator was used to take 3 measurements with each device at differing temperatures (22, 40, 45, 50, and 55 °C) and BP thresholds: 80/50, 100/60, 120/80, 140/90, 160/110, and 180/130 mm Hg. Using each device as its own control (22 °C), we determined the relative differences in mean BP for each device at each temperature and BP setting, assessed graphical trends with increasing temperature, and examined variability.
Results:
Graphical trends of mean simulator-subtracted BP differences from room temperature showed no discernable pattern, with differences clustered around zero. Overall mean difference in BP (combined elevated temperatures minus room temperature) was -0.8 ± 2.1 (systolic ± SD)/1.2 ± 3.5 (diastolic ± SD) mm Hg for the A&D device and 0.2 ± 0.4 (systolic ± SD)/-0.1 ± 0.1 (diastolic ± SD) mm Hg for the Omron. All individual elevated temperature differences (elevated temperature minus room temperature) except A&D diastolic BP at 50 °C were within 5 mm Hg.
Conclusions:
In this simulator-based study assessing within-device differences, higher ambient temperatures resulted in oscillometric BP measurements that were comparable to those performed at room temperature.
More Related Videos
14:09Tilt Testing with Combined Lower Body Negative Pressure: a "Gold Standard" for Measuring Orthostatic Tolerance
Published on: March 21, 2013
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Errors occurring during blood pressure monitoring
Several factors...
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)