Lead exposure in clinical imaging - The elephant in the room
Katia Katsari1, Chryssa Paraskevopoulou1, Milan Barati2
1Medical Diagnostic Imaging Department, College of Health Sciences, University of Sharjah, Sharjah 27272, United Arab Emirates.
Purpose:
Radiation Protection Apparel (RPA) is used during diagnostic imaging and interventional procedures to minimise incidental radiation exposure. The majority of RPA in use are lead-containing, which has until now been considered safe. A recent single-centre study has demonstrated that the external surface of 63 % of RPA was contaminated with lead dust. The purpose of this study was to reproduce this investigation with a larger sample size across Europe and assess whether decontamination procedures were possible.
Method:
The routine RPA Quality Control (QC) process was adapted to include lead dust contamination tests and decontamination if present. The presence of lead dust was determined using a commercially available colorimetric method. RPA that failed initial QC or could not be decontaminated were removed from use.
Results:
From June to October 2019, 728 RPA from 85 imaging centres from five countries underwent initial QC. Of these, 712 were tested for lead dust contamination which was present on 162 (23 %). Following cleaning, 85 (12 %) remained contaminated and were removed from use. Linear regression analysis shows a significant correlation between type of RPA and contamination, (p = 0.0015). There was no correlation between contamination and imaging department, year of manufacture, country and RPA condition (p-values 0.98, 0.90, 0.94 and 0.14 respectively).
Conclusions:
Lead dust contamination is present on 23 % of RPA that would pass routine QC procedures. Approximately half were not amenable to decontamination and were removed from use. Procedures were introduced for the routine handling of RPA, and updated QC steps for assessment and cleaning. Lead-free RPA should be considered.
More Related Videos
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
06:52Positron Emission Tomography Using 64-Copper as a Tracer for the Study of Copper-Related Disorders
Published on: April 28, 2023
Related Concept Videos
Radiological Investigation I: X-ray and CT
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies for Cardiovascular System V: CT