Wearable Cuff-Less Blood Pressure Estimation at Home via Pulse Transit Time
Objective:
We developed a wearable watch-based device to provide noninvasive, cuff-less blood pressure (BP) estimation in an at-home setting.
Methods:
The watch measures single-lead electrocardiogram (ECG), tri-axial seismocardiogram (SCG), and multi-wavelength photoplethysmogram (PPG) signals to compute the pulse transit time (PTT), allowing for BP estimation. We sent our custom watch device and an oscillometric BP cuff home with 21 healthy subjects, and captured the natural variability in BP over the course of a 24-hour period.
Results:
After calibration, our Pearson correlation coefficient (PCC) of 0.69 and root-mean-square-error (RMSE) of 2.72 mmHg suggest that noninvasive PTT measurements correlate with around-the-clock BP. Using a novel two-point calibration method, we achieved a RMSE of 3.86 mmHg. We further demonstrated the potential of a semi-globalized adaptive model to reduce calibration requirements.
Conclusion:
This is, to the best of our knowledge, the first time that BP has been comprehensively estimated noninvasively using PTT in an at-home setting. We showed a more convenient method for obtaining ambulatory BP than through the use of the standard oscillometric cuff. We presented new calibration methods for BP estimation using fewer calibration points that are more practical for a real-world scenario.
Significance:
A custom watch (SeismoWatch) capable of taking multiple BP measurements enables reliable remote monitoring of daily BP and paves the way towards convenient hypertension screening and management, which can potentially reduce hospitalizations and improve quality of life.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.