Related Experiment Video
Updated: Dec 9, 2025

Creating Rapid Oxygen Oscillations in Microbial Single-cell Growth Analysis using a Microfluidic Double-layer Device
Published on: July 18, 2025
Performance of mediator-less double chamber microbial fuel cell-based biosensor for measuring biological chemical
Minh Hang Do1, Huu Hao Ngo2, Wenshan Guo1
1Centre for Technology in Water and Wastewater, School of Civil and Environmental Engineering, University of Technology Sydney, Sydney, NWS, 2007, Australia.
Abstract:
Recently, the microbial fuel cell-based biosensor has been considered as an attractive technology for measuring wastewater quality such as biochemical oxygen demand (BOD). In this study, a mediator-less double compartment MFC based biosensor utilizing carbon felt as an anode electrode and inoculated with mixed culture was developed to improve the real application of a rapid BOD detection. This study aims to: (i) establish the effect of the operating conditions (i.e., pH, external resistance, fuel feeding rate) on MFC performance; (ii) investigate the correlation between biochemical oxygen demand (BOD) and signal output, and (iii) evaluate the operational stability of the biosensor. The presented result reveals that the maximum current and power production was obtained while 100 mM NaCl and 50 mM Phosphate buffer saline was used as a catholyte solution, neutral pH condition of media and fuel feeding rate at 0.3 mL min-1. Notably, a wider range of BOD concentration up to 300 mg L -1 can be obtained with the voltage output (R2 > 0.9901). Stable and steady power was produced by running MFC in 30 days when cells operated at 1000 Ω external resistance. Our research has some competition with the previous double chamber MFC in the upper limit of BOD detection. This results might help to increase the real application of MFC based BOD biosensor in real-time measurement.
More Related Videos
10:23Characterizing Mediated Extracellular Electron Transfer in Lactic Acid Bacteria with a Three-Electrode, Two-Chamber Bioelectrochemical System
Published on: August 23, 2024
08:31Anaerobic Protein Purification and Kinetic Analysis via Oxygen Electrode for Studying DesB Dioxygenase Activity and Inhibition
Published on: October 3, 2018