Enrichment of Oxygen Concentration Over Simulated Corneal Surface Through Noncontact Oxygen Delivery Device
Purpose:
To demonstrate a noncontact device to enrich oxygen concentration during corneal cross-linking (CXL).
Methods:
An oxygen delivery device was tested in a laboratory mock-up. The device comprises a clear polycarbonate tube of 14 cm in length and 1.58 cm inner diameter. Compressed oxygen gas is delivered to the tube from a side opening. The oximeter was attached to a sampling tube 3 mm above the apex of a scleral lens that simulates the cornea. The lens was mounted on a mannequin face. During each experimental run, the oximeter reading was recorded manually every 30 seconds for 4.5 minutes after the flow regulator was opened to the preset flow rate. Three flow rates of 0.25, 0.50, and 1 L/min were tested with all three cornea-tube distances of 8, 10, and 14 mm.
Results:
The baseline oxygen concentration was 20.9%. The oxygen concentration reached plateau levels after 2 to 3.5 minutes. Oxygen measurements were averaged over the three time points in the plateau phase between 3.5 and 4.5 minutes. Atmospheric oxygen concentration above the simulated cornea was found to be strongly dependent on the oxygen flow rate up to 1 L/min. At the 1 L/min flow rate, 99% concentration was achieved at 8 to 10 mm of cornea-tube distances, and dropped to 90% at 14 mm.
Conclusions:
Atmospheric oxygen concentration can be boosted to more than 90% using a noncontact device. This could potentially improve the effectiveness of accelerated CXL by boosting oxygen transport more than fourfold. [J Refract Surg. 2020;36(9):613-616.].
More Related Videos
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Oxygen Transport in the Blood
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...