False-Negative Serum Cryptococcal Antigen Lateral Flow Immunoassay Result for a Patient with Disseminated

Yuqiao Xu1,2, Wenying Xia1,2, Fang Ni1,2

  • 1Department of Laboratory Medicine, Jiangsu Province Hospital and Nanjing Medical University First Affiliated Hospital, Nanjing, People's Republic of China.

Insights

A false-negative cryptococcal antigen (CrAg) lateral flow assay (LFA) occurred despite a positive blood culture for Cryptococcus neoformans. This case highlights the postzone phenomenon, a potential pitfall in diagnosing fungal infections.

Area of Science:

  • Medical Mycology
  • Infectious Diseases
  • Diagnostic Immunology

Background:

  • Cryptococcus neoformans is an opportunistic fungal pathogen.
  • The polysaccharide capsule of C. neoformans is a key virulence factor and a target for diagnostics.
  • The cryptococcal antigen (CrAg) lateral flow assay (LFA) is widely used for diagnosis.