Accelerating Brain 3D T1-Weighted Turbo Field Echo MRI Using Compressed Sensing-Sensitivity Encoding (CS-SENSE)
Yunyun Duan1, Jie Zhang1, Zhizheng Zhuo1
1Department of Radiology, Beijing Tiantan Hospital, Capital Medical University, Beijing, China; Tiantan Image Research Center, China National Clinical Research Center for Neurological Diseases, Beijing, China.
Purpose:
To evaluate the clinical application of the accelerated 3D T1-weighted turbo field echo (T1W-TFE) using the compressed sensing-sensitivity encoding (CS-SENSE) and identify the appropriate acceleration factor.
Methods:
33 healthy controls (HC), 10 multiple sclerosis (MS) and 10 Alzheimer's disease (AD) patients were prospectively recruited. A conventional 3D T1W-TFE sequence and accelerated sequences with CS-SENSE factors of 3, 4.5, 6 and with SENSE factors of 3, 4.5 were acquired for all participants on a 3.0T MR system. The visual evaluation was independently assessed by two experienced radiologists. Quantitative image quality metrics and intraclass correlation coefficients (ICCs) between the conventional and the accelerated sequences were performed at the voxel level. Group comparisons were performed between HC and AD or MS patients.
Results:
There were no significant differences in the visual image quality metrics between conventional sequence and CS-SENSE factor of 3. The sequences with CS-SENSE factor of 6 and SENSE factors of 3, 4.5 showed significantly decreased overall image quality. The ICC values based on the voxel level of each accelerated scan and conventional scan were high (>0.9, 85.2%). For different accelerated sequences, AD and MS patients showed consistent results with the conventional sequence in gray matter atrophy when compared to HC.
Conclusions:
CS-SENSE factor of 3 is the appropriate parameter to accelerate the 3D T1W-TFE (65% time reduction) with preserved visual image quality. The voxel-based analysis demonstrated high ICCs for brain volume measurements in the majority of brain regions, implying the feasibility of the accelerated technique.
More Related Videos
10:143D Scanning Technology Bridging Microcircuits and Macroscale Brain Images in 3D Novel Embedding Overlapping Protocol
Published on: May 12, 2019
15:48Tracking the Mammary Architectural Features and Detecting Breast Cancer with Magnetic Resonance Diffusion Tensor Imaging
Published on: December 15, 2014
Related Concept Videos
Magnetic Resonance Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography