Related Experiment Video
Updated: Dec 9, 2025

Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
A Comparative Evaluation of Visual, Refractive, and Patient-Reported Outcomes of Three Extended Depth of Focus (EDOF)
Robert Edward Ang1,2, Gian Carlo S Picache2, Mark Christian R Rivera1
1Asian Eye Institute, Rockwell Center, Makati, Philippines.
Objective:
To compare visual, refractive, and patient-reported outcomes of patients implanted with one of three types of extended depth of focus (EDOF) intraocular lenses.
Setting:
Asian Eye Institute, Philippines.
Design:
Retrospective cohort study.
Methods:
Subjects implanted with Symfony (Johnson and Johnson, USA), IC-8 (AcuFocus, USA), and WIOL (Medicem, Czech Republic) EDOF intraocular lenses were recruited. Spherical equivalent, uncorrected and corrected visual acuity, defocus curve, and modulation-transfer function, Strehl ratio, and visual Strehl optical transfer function values, photic phenomena, and questionnaire answers were measured and assessed.
Results:
A total of 32 eyes with the Symfony lens, 30 with the IC-8 lens, and 32 with the WIOL lens were included in the study. Mean postoperative spherical equivalent was -0.24 D for the Symfony, -0.17 D for the IC-8, and 0.27 D for WIOL. There were no significant differences in postoperative monocular and binocular uncorrected and corrected visual acuity. On a monocular defocus curve, the IC-8 and Symfony showed significantly better vision than WIOL. The Symfony had significantly better modulation-transfer function, Strehl ratio, and visual Strehl optical transfer function. No difference was seen among the three lenses with regard to glare or starburst, while patient satisfaction remained high in all groups for far, intermediate, and near vision.
Conclusion:
All eyes implanted with the three EDOF designs achieved excellent far and intermediate vision, with acceptable near vision. The IC-8 and Symfony exhibited a better range of vision on defocus-curve testing. The Symfony showed superior results in quality of vision. Patient satisfaction was high in all three EDOF groups.
More Related Videos
06:55Scanning Light Scattering Profiler SLPS Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
Related Concept Videos
Focusing of Light in the Eye
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment