Related Experiment Video
Updated: Dec 9, 2025

Measuring Spatially- and Directionally-varying Light Scattering from Biological Material
Published on: May 20, 2013
Spatial scanning of a sample with two-dimensional angle-resolved low-coherence interferometry for analysis of
Ge Song1, Zachary A Steelman1, Wesley Kendall1
1Department of Biomedical Engineering, Duke University, Durham, NC 27708, USA.
Abstract:
Angle-resolved low-coherence interferometry (a/LCI) measures depth-resolved angular scattering for cell nuclear morphology analysis. 2D a/LCI, developed to collect across two scattering planes, is currently limited by the lack of spatial scanning. Here we demonstrate 2D a/LCI scanning across a three-dimensional volume using an image rotation scheme and a scanning mirror. Validation using various optical phantoms demonstrated excellent scatterer size determination over a 7.5 mm linear range, for a total accessible area of ∼44 mm2. Measurements from anisotropic scatterers allowed accurate determination of sizes and computation of aspect ratios. This scanning system will facilitate analysis of scatterer structure across wider tissue areas.