Related Experiment Video
Updated: Dec 8, 2025

MRI-guided Disruption of the Blood-brain Barrier using Transcranial Focused Ultrasound in a Rat Model
Published on: March 13, 2012
Focused ultrasound phantom model for blood brain barrier disruption
Tereza Alekou1, Marinos Giannakou1, Christakis Damianou1
1Department of Electrical Engineering and Computer Engineering and Informatics, Cyprus University of Technology, Cyprus.
Abstract:
In this paper a high intensity focused ultrasound (FUS) phantom model was developed, in order to be used in experiments for Blood Brain Barrier (BBB) disruption. The target was to create a phantom model that represents the disruption of the BBB during ultrasound application. An appropriate experimental setup was created bearing a single element transducer with diameter 50 mm and geometric focus 100 mm operating at 0.5 MHz. It included a set of tubes and a connector with multiple 0.4 mm openings, through which a suitable liquid is being circulated with a pump. The lesions were sealed with a thin homogenous layer of wax, preventing a liquid leakage. The system was tested successfully with FUS and a liquid leakage was achieved after FUS application. This set up is the first phantom model that has the potential to be utilized as a cost-effective solution for performing experiments for BBB disruption, without the need of using animal models.
More Related Videos
08:58A Benchtop Approach to the Location Specific Blood Brain Barrier Opening using Focused Ultrasound in a Rat Model
Published on: June 13, 2020
08:02A High-Throughput Image-Guided Stereotactic Neuronavigation and Focused Ultrasound System for Blood-Brain Barrier Opening in Rodents
Published on: July 16, 2020