Neglected lead tip erosion: An unusual case of S-ICD inappropriate shock
Gianfranco Mitacchione1, Marco Schiavone1, Alessio Gasperetti1
1Department of Cardiology, Luigi Sacco University Hospital, Milan, Italy.
Abstract:
A 52-year-old man experienced a subcutaneous implantable cardioverter-defibrillator (S-ICD) inappropriate shock due to electrode tip decubitus. The device, implanted two years before with a three-incision technique, was extracted, and a new electrode was implanted along the contralateral parasternal line with a two-incision technique, in a one-stage procedure. One-year follow-up was eventless. Early S-ICD electrode extraction and reimplantation during the same procedure is effective and should be considered as soon as initial signs of decubitus appear to avoid inappropriate shocks. A two-incision technique should be preferred to reduce the risk of electrode tip decubitus.
More Related Videos
Related Concept Videos
Acute Coronary Syndrome I: Introduction
Cardiopulmonary Resuscitation IV: Pharmacological Management
Blood Pressure Imbalances and Circulatory Shock
Blood Pressure: Hypertension and Hypotension
Normal blood pressure is 120/80 mm Hg. Elevated blood pressure is 120-129/under 80 mm Hg. Hypertension, warranting treatment at 130/80 mm Hg, is often asymptomatic and can lead to severe cardiovascular events, aneurysms, peripheral arterial disease, chronic renal disease, or cardiac...
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiopulmonary Resuscitation III: AED Use
Endocarditis IV: Nursing Management