Related Experiment Video
Updated: Dec 8, 2025

Guidelines for Elective Pediatric Fiberoptic Intubation
Published on: January 17, 2011
A comparison of videolaryngoscopy using standard blades or non-standard blades in children in the Paediatric
James Peyton1, Raymond Park1, Steven J Staffa2
1Department of Anesthesiology, Critical Care and Pain Medicine, Boston Children's Hospital, Boston, MA, USA; Department of Anaesthesia, Harvard Medical School, Boston, MA, USA.
Background:
The design of a videolaryngoscope blade may affect its efficacy. We classified videolaryngoscope blades as standard and non-standard shapes to compare their efficacy performing tracheal intubation in children enrolled in the Paediatric Difficult Intubation Registry.
Methods:
Cases entered in the Registry from March 2017 to January 2020 were analysed. We compared the success rates of initial and eventual tracheal intubation, complications, and technical difficulties between the two groups and by weight stratification.
Results:
Videolaryngoscopy was used in 1313 patients. Standard and non-standard blades were used in 529 and 740 patients, respectively. Both types were used in 44 patients. In children weighing <5 kg, standard blades had significantly greater success than non-standard blades at initial (51% vs 26%, P=0.002) and eventual (81% vs 58%, P=0.002) attempts at tracheal intubation. In multivariable logistic regression analysis, standard blades had 3-fold greater odds of success at initial tracheal intubations compared with non-standard blades (adjusted odds ratio 3.0, 95% confidence interval): 1.32-6.86, P=0.0009). Standard blades had 2.6-fold greater odds of success at eventual tracheal intubation compared with non-standard blades in children weighing <5 kg (adjusted odds ratio 2.6, 95% confidence interval: 1.08-6.25, P=0.033). There was no significant difference found in children weighing ≥5 kg.
Conclusions:
In infants weighing <5 kg, videolaryngoscopy with standard blades was associated with a significantly greater success rate than videolaryngoscopy with non-standard blades. Videolaryngoscopy with a standard blade is a sensible choice for tracheal intubation in children who weigh <5 kg.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Suctioning the Nasopharyngeal Airway
Equipment Required
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....