In vitro correlation between the effective and geometric orifice area in aortic stenosis
Jérôme Adda1, Viktoria Stanova2, Gilbert Habib3
1Department of Cardiology, Polyclinique les Fleurs, 83190 Ollioules, France.
Background:
Planimetry of aortic stenosis can be performed when Doppler measurements are unavailable. We sought to evaluate if, as advised in guidelines, the geometric orifice area (GOA) threshold value of 1 cm² was concordant with the threshold of 1 cm² of the effective orifice area (EOA), and the factors influencing the contraction coefficient (EOA/GOA ratio).
Methods:
In an in vitro mock circulatory system, we tested 6 degrees of AS severity (3 severe and 3 non-severe), and 3 levels of flow (<150 ml/s, 150-200 ml/s, >250 ml/s). The EOA was calculated by Doppler-echocardiography, and the GOA was measured with dedicated software after camera acquisition.
Results:
In all but the very low flow condition, an EOA of 1 cm² corresponded to a GOA of 1.2 cm². The contraction coefficient increased with both the flow and the stenosis severity. For very severe stenoses, the EOA and the GOA were interchangeable.
Conclusion:
As observed in clinical studies, the GOA was larger than the EOA, and a GOA between 1 and 1.2 cm² should not discard the possibility of severe aortic stenosis.