[Relationship between CT Numbers and Artifacts Obtained Using CT-based Attenuation Correction of PET/CT]
Yuya Watanabe1,2, Shota Hosokawa1, Yutaka Otaka3
1Graduate School of Health Sciences, Hirosaki University.
Computed tomography (CT)-based attenuation correction (CTAC) errors, caused by image misregistration, create artifacts in positron emission tomography (PET) non-activity signals. Careful CTAC is crucial to avoid inaccurate PET values and ensure reliable imaging results.
Area of Science:
- Medical Imaging
- Nuclear Medicine
- Radiology
Background:
- Computed tomography (CT)-based attenuation correction (CTAC) is essential for accurate positron emission tomography (PET) imaging.
- Image misregistration between CT and PET can lead to artifacts, particularly in regions with no radioactive tracer uptake (non-activity signals).
- Understanding these artifacts is critical for reliable quantitative analysis in PET scans.
Purpose of the Study:
- To investigate and characterize artifacts arising from CTAC errors due to image misregistration.
- To clarify the impact of varying CT numbers on the non-activity signal in PET images after CTAC.
Main Methods:
- A cylindrical phantom with a central tube served as the non-activity signal region.
- PET images were acquired using 18F-fluoro-2-deoxy-d-glucose (18F-FDG) in the background and water in the non-activity region.
- CT scans were performed with contrast agents to simulate various CT numbers, and PET images were corrected using these altered CT datasets.
Main Results:
- A direct correlation was observed between increasing CT numbers in the CT images and the magnitude of CTAC artifacts in PET.
- Distinct linear correlations were established for CTAC performed with CT numbers above zero (water) and below zero (air).
- Significant positive artifacts were detected in PET images post-CTAC, especially when CT numbers deviated from the original values (e.g., air).
Conclusions:
- Discrepancies in CT numbers used for CTAC, compared to the original scan, alter PET values.
- Image misregistration during CTAC can introduce significant artifacts, impacting quantitative accuracy.
- CTAC procedures require careful execution and validation to mitigate potential misregistration errors and ensure diagnostic integrity.
More Related Videos
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
09:21Human Brown Adipose Tissue Depots Automatically Segmented by Positron Emission Tomography/Computed Tomography and Registered Magnetic Resonance Images
Published on: February 18, 2015
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies for Cardiovascular System V: CT