Related Experiment Video
Updated: Dec 8, 2025

04:41
Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
527
A safe, stable, and convenient three-dimensional device for high Le Fort I osteotomy
Keisuke Sugahara1,2, Masahide Koyachi1, Kento Odaka3
1Department of Oral Pathobiological Science and Surgery, Tokyo Dental College, 2-9-18 Kanda Misaki-cho, Chiyoda-ku, Tokyo, Japan.
Maxillofacial Plastic and Reconstructive Surgery
|September 23, 2020
Summary
A novel 3D device combined with piezosurgery offers a safe, inexpensive method for high Le Fort I osteotomy. This technique prevents tooth root injury and improves surgical field access for patients with short midfaces.
Area of Science:
- Oral and Maxillofacial Surgery
- Orthognathic Surgery
- 3D Technology in Medicine
Background:
- Le Fort I osteotomy is a standard procedure for skeletal jaw deformities.
- High Le Fort I osteotomy elevates the osteotomy site to address midface hypoplasia.
- Advancements in 3D printing have increased its application in orthognathic surgery.
Purpose of the Study:
- To present a novel, cost-effective method for high Le Fort I osteotomy.
- To demonstrate the use of a new 3D device and piezosurgery for improved surgical outcomes.
- To address challenges associated with conventional 3D devices in specific patient cases.
Main Methods:
- A modified 3D device was designed in two separate parts for better surgical field access.
- The 3D device was stabilized using fixation points at the anterior nasal spine and piriform aperture.
- Piezosurgery was employed to ensure precise osteotomy and prevent damage to vital structures.
Main Results:
- The novel 3D device facilitated accurate high Le Fort I osteotomy in a patient with a short midface and long tooth roots.
- The surgical technique successfully prevented injury to tooth roots and the infraorbital nerve.
- The thin, custom-designed 3D device did not interfere with the surgical field, enhancing safety and accuracy.
Conclusions:
- The developed 3D device is smaller, simpler, and more cost-effective than conventional devices.
- This method provides an efficient solution for accurate high Le Fort I osteotomy.
- The combined use of the novel 3D device and piezosurgery ensures safe and precise surgical execution.