Design and Characterization of a Minimally Invasive Bipolar Electrode for Electroporation
Giulia Merola1, Roberta Fusco1, Elio Di Bernardo1
1Oncology Medical and Research & Development Division, Igea SpA, 41012 Carpi, Italy.
Objective:
To test a new bipolar electrode for electroporation consisting of a single minimally invasive needle.
Methods:
A theoretical study was performed by using Comsol Multiphysics® software. The prototypes of electrode have been tested on potatoes and pigs, adopting an irreversible electroporation protocol. Different applied voltages and different geometries of bipolar electrode prototype have been evaluated.
Results:
Simulations and pre-clinical tests have shown that the volume of ablated area is mainly influenced by applied voltage, while the diameter of the electrode had a lesser impact, making the goal of minimal-invasiveness possible. The conductive pole's length determined an increase of electroporated volume, while the insulated pole length inversely affects the electroporated volume size and shape; when the insulated pole length decreases, a more regular shape of the electric field is obtained. Moreover, the geometry of the electrode determined a different shape of the electroporated volume. A parenchymal damage in the liver of pigs due to irreversible electroporation protocol was observed.
Conclusion:
The minimally invasive bipolar electrode is able to treat an electroporated volume of about 10 mm in diameter by using a single-needle electrode. Moreover, the geometry and the electric characteristics can be selected to produce ellipsoidal ablation volumes.
More Related Videos
10:34The Fabrication and Operation of a Continuous Flow, Micro-Electroporation System with Permeabilization Detection
Published on: January 7, 2022
09:29Preparation and Implantation of Electrodes for Electrically Kindling VGAT-Cre Mice to Generate a Model for Temporal Lobe Epilepsy
Published on: August 17, 2021