Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Agreement and repeatability of Icare ic100 tonometer
Judy Jose1, Ramesh S Ve1, H Vijaya Pai2
1Department of Optometry, Manipal College of Health Professions, Manipal Academy of Higher Education (MAHE), Manipal, Karnataka, India.
The Icare ic100 tonometer demonstrates good reliability and repeatability for measuring intraocular pressure (IOP). This rebound tonometer offers accurate IOP readings comparable to Goldmann applanation tonometry across various pressure ranges.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Medical Devices
Background:
- Intraocular pressure (IOP) measurement is crucial for diagnosing and managing glaucoma.
- Accurate and repeatable tonometry is essential for reliable IOP assessment.
- The Icare ic100 is a non-contact rebound tonometer.
Purpose of the Study:
- To evaluate the agreement and repeatability of the Icare ic100 tonometer.
- To compare the Icare ic100 with the Icare TA01i and Goldmann applanation tonometer (GAT).
Main Methods:
- A cross-sectional, multicenter study included 150 adult subjects with IOP ≥7 mmHg.
- Two masked examiners performed five IOP measurements using Icare ic100, Icare TA01i, and GAT.
- Intraclass correlation coefficient (ICC) and Cronbach's alpha quantified agreement; repeatability coefficient (RC) and coefficient of variation (CV) assessed test-retest variability.
Main Results:
- The Icare ic100 showed good reliability (ICC >0.78) compared to the Icare TA01i across all IOP groups.
- Compared to GAT, the Icare ic100 demonstrated good reliability (ICC >0.87), with moderate reliability (ICC = 0.52) in the >16 to <23 mmHg IOP range.
- The Icare ic100 exhibited good repeatability with RC of 2.67 and CV of 4.89.
Conclusions:
- The Icare ic100 rebound tonometer provides reliable and repeatable IOP measurements with minimal error.
- It demonstrates comparable performance to GAT across a wide IOP range.
- The Icare ic100 is a suitable device for IOP assessment without compromising corneal health.
Related Concept Videos
Uncertainty in Measurement: Reading Instruments
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessment of blood pressure in brachial artery(two-step method)
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: