A Protocol to Set Up Needle-Free Connector with Positive Displacement on Central Venous Catheter in Intensive Care Unit
Published on: July 13, 2019
Manual Shunt Connector Tool to Aid in No-Touch Technique
Grace Y Lai1,2, Pascal Voyer-Nguyen2, Thomas Looi2
1Department of Neurological Surgery , McGaw Medical Center of Northwestern University, Chicago, Illinois.
Background:
Given the morbidity and cost associated with cerebrospinal fluid shunt infections, many neurosurgical protocols implement "no-touch" technique to minimize infection. However, current surgical tools are not designed specifically for this task and surgeons often resort to using their hands to connect the shunt catheter to the valve.
Objective:
To develop an efficient and effective shunt assembly tool.
Methods:
Prototypes were designed using computer assisted software and machined in stainless steel. The amount of time and number of attempts it took volunteers to connect a Bacticel shunt catheter to a Delta valve were recorded using the new tool and standard shodded mosquitos. Scanning electron microscopy (SEM) was done on manipulated catheters to assess potential damage. Practicing neurosurgeons provided feedback.
Results:
Nonsurgeon (n = 13) volunteers and neurosurgeons (n = 6) both completed the task faster and with fewer attempts with the new tool (mean 7.18 vs 15.72 s and 2.00 vs 6.36 attempts, P < .0001; mean 2.93 vs 5.96 s and 1.06 vs 2.94 attempts, P < .001, respectively). SEM of 24 manipulated catheters showed no microscopic damage. 100% of neurosurgeons surveyed (n = 10) would adapt the tool in their practice, 90% preferred use of the new tool compared to their existing method, and 100% rated it easier to use compared to existing instruments.
Conclusion:
The new tool shortened the time and number of attempts to connect a shunt catheter to a valve. Neurosurgeons preferred the new tool to existing instruments. There was no evidence of catheter damage with the use of this tool.