Deep learning-based synthetic CT generation for paediatric brain MR-only photon and proton radiotherapy
Matteo Maspero1, Laura G Bentvelzen1, Mark H F Savenije1
1Department of Radiotherapy, Division of Imaging & Oncology, University Medical Center Utrecht, , The Netherlands; Computational Imaging Group for MR diagnostics & therapy, Center for Image Sciences, University Medical Center Utrecht, , The Netherlands.
Background And Purpose:
To enable accurate magnetic resonance imaging (MRI)-based dose calculations, synthetic computed tomography (sCT) images need to be generated. We aim at assessing the feasibility of dose calculations from MRI acquired with a heterogeneous set of imaging protocol for paediatric patients affected by brain tumours.
Materials And Methods:
Sixty paediatric patients undergoing brain radiotherapy were included. MR imaging protocols varied among patients, and data heterogeneity was maintained in train/validation/test sets. Three 2D conditional generative adversarial networks (cGANs) were trained to generate sCT from T1-weighted MRI, considering the three orthogonal planes and its combination (multi-plane sCT). For each patient, median and standard deviation (σ) of the three views were calculated, obtaining a combined sCT and a proxy for uncertainty map, respectively. The sCTs were evaluated against the planning CT in terms of image similarity and accuracy for photon and proton dose calculations.
Results:
A mean absolute error of 61 ± 14 HU (mean±1σ) was obtained in the intersection of the body contours between CT and sCT. The combined multi-plane sCTs performed better than sCTs from any single plane. Uncertainty maps highlighted that multi-plane sCTs differed at the body contours and air cavities. A dose difference of -0.1 ± 0.3% and 0.1 ± 0.4% was obtained on the D > 90% of the prescribed dose and mean γ2%,2mm pass-rate of 99.5 ± 0.8% and 99.2 ± 1.1% for photon and proton planning, respectively.
Conclusion:
Accurate MR-based dose calculation using a combination of three orthogonal planes for sCT generation is feasible for paediatric brain cancer patients, even when training on a heterogeneous dataset.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
10:48PET and MRI Guided Irradiation of a Glioblastoma Rat Model Using a Micro-irradiator
Published on: December 28, 2017
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...