[Investigation of Successful Eyedrop Instillation Rates and Analysis of Drop Positions Using High-speed Digital Video
Hiroaki Ikeda1, Ayane Takamoto1, Junko Ikeda2
1Department of Pharmaceutical Health Care and Sciences, Kagawa School of Pharmaceutical Sciences, Tokushima Bunri University.
Yakugaku Zasshi : Journal of the Pharmaceutical Society of Japan
|October 1, 2020
Summary
Eyedrop instillation success rates were evaluated. Most patients achieved a 70.1% success rate on the first attempt, with the ideal dropper-to-cornea distance being approximately 1.6 cm for optimal results.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Patient Adherence Studies
Background:
- Eyedrop adherence studies often assume successful instillation.
- The optimal distance between the eyedropper and the cornea is not well-defined, with a common estimate of 2.54 cm (1 inch).
- Understanding accurate eyedrop instillation techniques is crucial for effective ocular treatment.
Purpose of the Study:
- To investigate the success rates of eyedrop instillation in a volunteer cohort.
- To determine the relationship between the distance from the dropper tip to the cornea and successful eyedrop delivery.
- To assess current eyedrop instillation practices and identify areas for improvement.
Main Methods:
- High-speed digital video recording was used to analyze eyedrop instillation in 100 volunteers.
- Successful instillation was defined as accurate delivery of one drop on the first attempt.
- Distances were measured from paused video frames using a digital ruler.
Main Results:
- The overall success rate for first-time eyedrop instillation was 70.1%.
- The average distance between the dropper tip and cornea was 2.62±1.75 cm (median 2.20 cm).
- Success rates varied with distance: 77% at 1.6±0.88 cm versus 54.9% at 4.8±1.25 cm.
Conclusions:
- The generally recommended eyedrop instillation distance is often followed, but not always optimal.
- A shorter distance (around 1.6 cm) between the dropper tip and cornea significantly improves successful eyedrop instillation rates.
- Patient education on proper eyedrop technique is essential, as 40% of volunteers received no prior instructions.