Comparison of Torque, Screw-in Force, and Shaping Ability of Glide Path Instruments in Continuous Rotation and
Pyae Hein Htun1, Arata Ebihara1, Keiichiro Maki1
1Department of Pulp Biology and Endodontics, Division of Oral Health Sciences, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University (TMDU), Tokyo, Japan.
Introduction:
This study aimed to analyze torque/force generation and transportation in double-curved canals instrumented with 3 types of glide path files using optimum glide path (OGP) motion in comparison with continuous rotation.
Methods:
Sixty simulated double-curved canals were prepared with #10/0.05 or #15/0.03 HyFlex EDM Glidepath files (Coltene/Whaledent, Altstätten, Switzerland) or a #13/0.04 prototype MANI Glidepath file (Tochigi, Japan) using OGP motion or continuous rotation (n = 10 each). Canals were sequentially prepared to 20 mm and 22 mm (full working length) using automated root canal instrumentation and a torque/force analyzing device. Transportation was calculated at 1-9 mm from the apex. Data were compared using 2-way analysis of variance followed by a post hoc simple main effect test with Bonferroni correction and a Kruskal-Wallis test (α = 5%).
Results:
All #10/0.05 instruments fractured. In the 22-mm preparation, the OGP motion resulted in lower clockwise torque and screw-in force than did continuous rotation (P < .05). In the 20-mm preparation, #15/0.03 instruments recorded a lower screw-in force for OGP motion than for continuous rotation (P < .05). Comparing the 2 preparation phases, OGP motion generated no significant differences; however, continuous rotation developed higher clockwise torque and screw-in force in the 22-mm preparation than in the 20-mm preparation (P < .05). There was no significant difference among the tested groups for transportation values.
Conclusions:
Compared with continuous rotation, OGP motion generated less screw-in force, lower clockwise torque, and similar transportation. The #15/0.03 HyFlex EDM instrument and the #13/0.04 prototype MANI instrument performed similarly well.
Related Concept Videos
Torque
Torque can be considered as the rotational counterpart to force. Since forces change the translational...
Torque Free Motion
Screw: Problem Solving
To evaluate the...
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Residual Stresses in Circular Shafts
Rotational Motion about a Fixed Axis