Related Experiment Video
Updated: Dec 7, 2025

Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Infolding of Self-Expandable Transcatheter Heart Valve: Case Report and Review of Literature
Wassef Karrowni1, Saleh Fakih2, Pierre Nassar3
1Cardiovascular Disease, UnityPoint Health, Cedar Rapids, USA.
Abstract:
Transcutaneous aortic valve replacement (TAVR) has become a widely accepted minimally invasive approach for treatment of severe aortic stenosis. Self-expandable prostheses are commonly the device of choice, with excellent procedural success and durability. However, there have been several recent case reports of infolding of the self-expandable prosthesis during development with subsequent malfunction and need for further intervention. We present a case of self-expandable valve prosthesis infolding managed by balloon postdilation, and summarize the cases reported in the literature to date in an attempt to increase awareness of this serious technical problem and the factors associated with it.
Related Concept Videos
Mitral Valve Prolapse II: Assessment and Management
Mitral Valve Prolapse I: Introduction
Mitral Stenosis III: Medical Management