Should we recommend occipital plate fixation using bicortical screws or inverted occipital hooks to optimize
Arnaud Germaneau1, Tanguy Vendeuvre1, Alexandre Delmotte2
1Institut Pprime UPR 3346, CNRS - Université de Poitiers - ISAE-ENSMA, France; Spine & Neuromodulation Functional Unit, Department of Neurosurgery, CHU Poitiers, PRISMATICS Lab, Poitiers, France.
Background:
Occipito-cervical fusion can be necessary in case of cranio-cervical junction instability. Proximal stabilisation is usually ensured by bi-cortical occipital screws implanted through one median or two lateral occipital plate(s). Bone thickness variability as well as the proximity of vasculo-nervous elements can induce substantial morbidity. The choice of site and implant type remains difficult for surgeons and is often empirically based. Given this challenge, implants with smaller pitch to increase bone interfacing are being developed, as is a surgical technique consisting in inverted occipital hook clamps, a potential alternative to plate/screws association. We present here a biomechanical comparison of the different occipito-cervical fusion devices.
Methods:
We have developed a 3D mark tracking technique to measure experimental mechanical data on implants and occipital bone. Biomechanical tests were performed to study the mechanical stiffness of the occipito-cervical instrumentation on human skulls. Four occipital implant systems were analysed: lateral plates+large pitch screws, lateral plates+hooks, lateral plates+small pitch screws and median plate+small pitch screws. Mechanical responses were analysed using 3D displacement field measurements from optical methods and compared with an analytical model.
Findings:
Paradoxical mechanical responses were observed among the four types of fixations. Lateral plates+small pitch screws appear to show the best accordance of displacement field between bone/implant/system interface providing higher stiffness and an average maximum moment around 50 N.m before fracture.
Interpretation:
Stability of occipito-cervical fixation depends not only on the site of screws implantation and occipital bone thickness but is also directly influenced by the type of occipital implant.
More Related Videos
12:25C-arm-Free Simultaneous OLIF51 and Percutaneous Pedicle Screw Fixation in a Single Lateral Position
Published on: September 16, 2022
09:07An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
Related Concept Videos
Sutures of the Skull
Sutures are immobile joints between adjacent bones of the skull. The narrow gap between the bones is filled with dense, fibrous connective tissue that unites the bones. The long sutures located between the skull bones are not straight but instead follow irregular, tightly twisting paths. These twisting lines tightly...
Cranial Bones: Superior and Posterior View
The frontal bone is the single bone that forms the forehead. At its anterior midline, between the eyebrows, there is a slight depression called the glabella. The frontal bone also forms the supraorbital margin of the orbit. Near the middle of this margin is the supraorbital foramen, the opening that provides passage for a sensory nerve to the forehead. The frontal bone is thickened just above each supraorbital margin,...