Current clinical methods of measurement of respiratory rate give imprecise values
Gordon B Drummond1, Darius Fischer2, D K Arvind2
1Dept of Anaesthesia, Critical Care, and Pain Medicine, University of Edinburgh, Edinburgh UK.
Background:
Respiratory rate is a basic clinical measurement used for illness assessment. Errors in measuring respiratory rate are attributed to observer and equipment problems. Previous studies commonly report rate differences ranging from 2 to 6 breaths·min-1 between observers.
Methods:
To study why repeated observations should vary so much, we conducted a virtual experiment, using continuous recordings of breathing from acutely ill patients. These records allowed each breathing cycle to be precisely timed. We made repeated random measures of respiratory rate using different sample durations of 30, 60 and 120 s. We express the variation in these repeated rate measurements for the different sample durations as the interquartile range of the values obtained for each subject. We predicted what values would be found if a single measure, taken from any patient, were repeated and inspected boundary values of 12, 20 or 25 breaths·min-1, used by the UK National Early Warning Score, for possible mis-scoring.
Results:
When the sample duration was nominally 30 s, the mean interquartile range of repeated estimates was 3.4 breaths·min-1. For the 60 s samples, the mean interquartile range was 3 breaths·min-1, and for the 120 s samples it was 2.5 breaths·min-1. Thus, repeat clinical counts of respiratory rate often differ by >3 breaths·min-1. For 30 s samples, up to 40% of National Early Warning Scores could be misclassified.
Conclusions:
Early warning scores will be unreliable when short sample durations are used to measure respiratory rate. Precision improves with longer sample duration, but this may be impractical unless better measurement methods are used.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Respiratory Volumes and Capacities I
Factors Affecting Respiration
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration...