Small changes in hand height alter absorbance, but not pulsation, in the finger pulse plethysmograph
Onur Howie1, Jamie R K Marland2, Ralph Shackleton3
1Edinburgh Medical School, University of Edinburgh, Chancellor's Building, 49 Little France Crescent, Edinburgh EH16 4SB, United Kingdom.
Abstract:
Objective.Previous studies of finger pulse plethysmography have used large changes in finger height to assess how pressure influences light transmission. Light transmission is affected by pulsation transmitted from the arterial pressure, and also by the volume of blood held in the finger vasculature. Smaller changes of hand height, in relation to the neck, could elucidate more precisely the pressures and mechanisms associated with vascular volume changes.Approach.We studied 18 (15 male, 3 female) healthy young volunteers in warm quiet surroundings, with one arm supported horizontally at the level of the shoulder. Comparisons were made with the hand placed 5 cm above and 5 cm below the horizontal, in random sequence. Red and infrared transmission were recorded from a standard pulse oximeter finger probe. Sections of recording were chosen where short periods of vasoconstriction were absent. Total absorbance values at peak and trough, and absorbance pulse amplitude, were calculated. We compared measures in the two hand positions, and before and after inflation of a cuff at the wrist, within each subject.Main results.In the dependent position, infrared light absorbances were greater: peak, 0.054 (0.043, 0.070) and trough (0.053 (0.045, 0.069) (median, quartiles) (P< 0.0001). Pulse amplitude was unaffected. Cuff inflation had no effect on the peak and trough values, but pulse amplitude decreased: the difference between paired values for infrared was 0.0049 (0.0075, 0.0015) and red, 0.0030 (0.0043, 0.0010) (median paired difference, 95% confidence limits). (P< 0.005).Significance.The minimal effect of cuff venous obstruction supports previous work reporting flow limitation in venous drainage of the arm. Our observations suggest that an increase in filling pressure caused by lowering the hand by 10 cm increases overall blood volume in the finger, suggesting that the plethysmograph waveform may result from pressure changes much less than arterial.
Related Concept Videos
Special considerations while measuring pulse
Pulse
The pulse serves as a clinical indicator...
Cardiovascular System Abnormal Findings I: Inspection and Palpation
Abnormal findings observed during an inspection
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: