Measuring the Carotid to Femoral Pulse Wave Velocity Cf-PWV to Evaluate Arterial Stiffness
Published on: May 3, 2018
Methodological considerations for the measurement of arterial stiffness using applanation tonometry
Alexandra B Cooke1, Alvin Kuate Defo2, Kaberi Dasgupta3,4
1Division of Experimental Medicine, Department of Medicine, McGill University.
Introduction:
Accurate comparisons of carotid--femoral pulse wave velocity (cfPWV) within and across studies require standardized procedures. Guidelines suggest reporting the average of at least two cfPWV measurements; if the difference exceeds 0.5 m/s, a third measurement should be taken, and the median reported. Another method involves repeating measurements until two values are within 0.5 m/s. However, in many studies, duplicate measurements are averaged irrespective of the difference between readings. We evaluated the impact of these methods on the reported cfPWV value.
Methods:
Measurements of cfPWV (SphygmoCor) from five studies included individuals spanning a wide age range, with or without comorbid conditions, and pregnant women. In participants with at least three high-quality measurements, differences between the median value (MED) and the average of the first two cfPWV measurements (AVG1) and the average of two cfPWV measurements within 0.5 m/s (AVG2) were evaluated using paired t-tests and Bland--Altman plots.
Results:
Participants' mean age was 50 ± 14 years and BMI was 28.0 ± 5.5 kg/m2 (N = 306, 79% women). The overall mean difference was -0.10 m/s (95% CI 0.17 to -0.04) between MED and AVG1, and 0.11 m/s (95% CI 0.05--0.17) between MED and AVG2. The absolute difference exceeded 0.5 m/s in 34% (MED-AVG1) and 22% (MED-AVG2) of participants, and 1 m/s in 8% of participants (both MED-AVG1 and MED-AVG2). Scatter around the bias line increased with higher mean cfPWV values.
Conclusion:
Although the overall mean difference in cfPWV between protocols was not clinically relevant, large variation led to absolute differences exceeding 0.5 m/s in a large proportion of participants.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure