Core temperatures (Tcore) are often invasive, and can be underutilized. Peripheral temperatures are easier to obtain, but are often less accurate. A zero-heat-flux thermometer (ZHF) is a non-invasive method to obtain core temperatures (TZHF), and has been accurate when compared to Tcore in the operating room. We aimed to determine whether TZHF accurately and reliably measures Tcore in emergency department (ED) patients when compared to rectal, bladder or esophageal temperatures.
Methods:
We conducted a prospective observational quality improvement project, with concurrent TZHF and Tcore measurements. The primary outcome was whether one device detected a fever (≥38.1 °C) when the other device did not. Unadjusted linear regression was used to determine the relationship between temperature differences between devices.
Results:
268 patients were included. Mean temperatures were 36.6 °C for Tcore and 36.3 °C for TZHF. 16 of 52 patients with fever identified by Tcore were not detected by TZHF, 13 with an infectious etiology. The mean temperature difference between Tcore and TZHF increased as the patient's temperature increased; the difference was 0.2 °C in afebrile patients, but 0.7 °C in febrile patients.
Conclusion:
While we found overall concordance between Tcore and TZHF, the ZHF did not detect fever in 25% of patients presenting with fever of infectious origin. Measurements between Tcore and TZHF varied more as temperatures increased, with TZHF consistently reporting lower values. Although more study is needed, these findings call into question the use of TZHF in the ED where detection of fever frequently guides patient evaluation and management.
More Related Videos
05:43Author Spotlight: Improving Lesion Contiguity in Pulmonary Vein Isolation via Proactive Esophageal Cooling
Published on: April 19, 2024
04:58Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...