DXA body composition corrective factors between Hologic Discovery models to conduct multicenter studies.
Thibault Sutter1, François Duboeuf2, Roland Chapurlat2
1EA 4708 I3MTO Laboratory, University of Orleans, Orleans, France.
Bone
|October 12, 2020
Summary
Dual X-ray absorptiometry (DXA) body composition measurements show significant inter-center variability. This study quantifies these differences, proposing corrective factors essential for accurate multicenter research using Hologic Discovery DXA scanners.
Area of Science:
- Biomedical Engineering
- Clinical Densitometry
- Body Composition Analysis
Background:
- Dual X-ray absorptiometry (DXA) is a standard for body composition assessment in clinical and research settings.
- Variability in DXA measurements across different instruments necessitates caution in longitudinal studies and multicenter research.
- Standardization is crucial for reliable interpretation of body composition data.
Purpose of the Study:
- To quantify intra- and inter-center variability of bone mineral content, bone mineral density, fat, and lean body composition measurements.
- To compare Hologic Discovery DXA models for consistency in body composition assessment.
- To calculate corrective factors for multicenter research projects using DXA.
Main Methods:
- A whole-body phantom was analyzed ten times across three centers using Hologic Discovery DXA scanners (QDR A and QDR W).
- Phantom analysis included bone, lean, and fat mass measurements.
- Stability and accuracy were monitored using vertebral and block phantoms.
Main Results:
- All devices demonstrated good long-term stability and accuracy.
- Intra-center coefficients of variation were within manufacturer-acceptable ranges (e.g., bone mineral density: 1.40%).
- Inter-center coefficients of variation exceeded 8% for bone mineral density and lean mass, highlighting significant discrepancies.
Conclusions:
- Significant inter-center discrepancies in DXA body composition measurements remain a major concern.
- Cross-calibration of DXA scanners is necessary for multicenter studies.
- Corrective factors derived from phantom studies are proposed to adjust multicenter DXA data.