Related Experiment Video
Updated: Dec 5, 2025

Tachycardia-Induced Cardiomyopathy As a Chronic Heart Failure Model in Swine
Published on: February 17, 2018
Pacemaker-mediated tachycardia in a dual-lead CRT-D: What is the mechanism?
Christopher Monkhouse1, Alex Cambridge1, Anthony W C Chow2
1Barts Heart Centre, West Smithfield, London, UK.
Abstract:
A 73-year-old gentleman with dilated cardiomyopathy, left bundle branch block and a left ventricular (LV) ejection fraction of 20% was implanted with two LV leads in a tri-ventricular cardiac resynchronisation therapy defibrillator (CRT-D) trial. As a part of the trial he was programmed with fusion-based CRT therapy with dual LV lead only pacing. The patient presented to local heart failure service 12 years after implant, after a positive response to CRT therapy, with increase in fatigue, shortness of breath and bilateral pitting oedema. The patient sent a remote monitoring transmission that suggested loss of capture on one of the LV leads. This coupled with atrial ectopics was producing a high burden of pacemaker-mediated tachycardia (PMT) that was not seen when both LV leads had been capturing. What is the mechanism for this? Dual LV-lead tri-ventricular leads have been shown to have variable improvements in CRT response but with an increased complexity of implant procedure. This is the first case report of PMT-induced heart failure exacerbation in a tri-ventricular device following loss of LV capture of one lead.
More Related Videos
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Dysrhythmias III: Characteristics of Dysrhythmias
Electrophysiology of Normal Cardiac Rhythm
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Conduction System of the Heart
This system relies on the unique properties of nodal and Purkinje cells:...
Mechanism of Cardiac Arrhythmias