Related Experiment Video
Updated: Dec 5, 2025

Use of a Multi-compartment Dynamic Single Enzyme Phantom for Studies of Hyperpolarized Magnetic Resonance Agents
Published on: April 15, 2016
MRI of [2-13 C]Lactate without J-coupling artifacts
Keshav Datta1, Daniel Spielman1
1Department of Radiology, Stanford University, Stanford, California, USA.
Purpose:
Imaging of [2-13 C]lactate, a metabolic product of [2-13 C]pyruvate, is over considerable interest in hyperpolarized 13 C studies. However, artifact-free imaging of a J-coupled nuclear spin species can be challenging due to the peak-splitting induced by the spin-spin interactions. In this work, two new techniques resolving these J-modulated artifacts are presented.
Theory And Methods:
The Product Operator Formalism (POF) of density matrix theory is used to both numerically and analytically derive the coherences arising during radiofrequency excitation and readout of a J-coupled spin system. A combination of computer simulations and experiments with [2-13 C]lactate and 13 C-formate phantoms are then used to verify the performance of two imaging methods. In the first approach, a quadrature imaging technique is used to eliminate scalar coupling artifacts via the combination of in-phase and quadrature images acquired at echo times differing by 1/2J with an echoplanar readout. The second approach employs a highly narrowband RF excitation pulse to image a single peak from the J-coupled doublet.
Results:
Simulations using a numerical Shepp-Logan phantom, in vitro experiments using thermally polarized [2-13 C]lactate, thermally and hyperpolarized 13 C-formate phantoms, and in vivo imaging of [2-13 C]lactate produced in rat brain following injection of hyperpolarized [2-13 C]pyruvate show artifact-free images and demonstrate potential utility of these methods.
Conclusion:
The quadrature imaging and the narrowband excitation techniques resolve the J-coupling induced ghosting and blurring artifacts present with conventional MRI of J-coupled signals such as [2-13 C]lactate.
More Related Videos
11:43Hyperpolarized 13C Metabolic Magnetic Resonance Spectroscopy and Imaging
Published on: December 30, 2016
13:56The Use of Magnetic Resonance Spectroscopy as a Tool for the Measurement of Bi-hemispheric Transcranial Electric Stimulation Effects on Primary Motor Cortex Metabolism
Published on: November 19, 2014
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies for Cardiovascular System IV: CMRI
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...
2D NMR: Overview of Homonuclear Correlation Techniques
COSY90 is the standard two-dimensional (2D) COSY experiment that...