Computed tomography arthrography of the shoulder with tin filter-based spectral shaping at 100 kV and 140 kV
Yun Seok Choi1, Hye Jung Choo1, Sun Joo Lee1
1Department of Radiology, Inje University Busan Paik Hospital, Busan, Republic of Korea.
Background:
Tin filter-based spectral shaping has been used for low-dose and ultra-low-dose computed tomography (CT) in several body parts. However, studies of shoulder CT arthrography with spectral shaping are limited.
Purpose:
To investigate image quality and radiation dose of shoulder CT arthrography with tin filter-based spectral shaping at 100 kV (Sn 100 kV) and 140 kV (Sn 140 kV) in comparison with the conventional protocol.
Material And Methods:
Ninety-nine shoulder CT arthrographies with protocols of Sn 100 kV (n = 32), Sn 140 kV (n = 25), and conventional 120 kV (n = 42) were retrospectively evaluated. Qualitative image quality, CT attenuations of intra-articular contrast mixture and tissues, background noise, contrast-to-noise ratios (CNRs), and figures of merit were assessed. Radiation doses were compared.
Results:
CT arthrographies with Sn 100 kV and Sn 140 kV yielded approximately 70% and 60% radiation dose reduction, respectively, compared with the conventional 120 kV (P < 0.001). Qualitative image noise and quantitative background noise of Sn 100 kV and Sn 140 kV were significantly less than those of the conventional protocol. Qualitative image contrast, CT attenuations of intra-articular contrast mixture and tissues, and CNRs for Sn 100 were similar to those of the conventional 120 kV. However, Sn 140 kV showed significantly lower qualitative contrast and CNRs than 120 kV. Sn 100 kV was the most dose efficient among the three protocols.
Conclusion:
Shoulder CT arthrography with Sn 100 kV substantially reduced radiation dose and image noise and maintained image contrast, compared with the conventional protocol.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...