Quantifying semi-volatile organic contaminants in solution by internal standard addition method requires prompt
1Environmental Monitoring Division, City of Los Angeles, Playa Del Rey, CA, 90293, USA. ochan.otim@lacity.org.
Abstract:
The precision and accuracy of quantifying semi-volatile organic compounds (SVOCs) in solution by GC/MS, particularly when volume errors are unpredictable or difficult to control, are improved by utilizing internal standards (IS). Not obvious though is the extent to which timing IS addition affects measurement. To illustrate this fact, the mean concentrations of 60 SVOCs (40 or 80 μg/mL) in two identical solutions into which IS were added at different times are compared in this study. In one solution, IS were added promptly on preparation (reference); in the other, IS were added after 36 days of incubation (treatment). To investigate the role that temperature might play here as well, equal fractions of each solution were incubated at - 20 °C, 4 °C or 22 °C. Results, as determined by one-way ANOVA, show that there were no differences between the reference solutions at the beginning and after 36 days (F3,236 = 0.244, p = 0.865), but that significant differences exist between the reference solutions collectively and the treatment irrespective of temperature (F6,413 = 6.76, p = 1.99e-06). These results, confirmed by a post hoc analysis, suggest that uncertainty is introduced into SVOC quantitation when internal standards are not added promptly into SVOCs solutions on preparation.
More Related Videos
04:56In Vitro Method to Control Concentrations of Halogenated Gases in Cultured Alveolar Epithelial Cells
Published on: October 23, 2018
13:16A Whole Cell Bioreporter Approach to Assess Transport and Bioavailability of Organic Contaminants in Water Unsaturated Systems
Published on: December 24, 2014
Related Concept Videos
Standard Solutions
Volatilization
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...