Gantry angle dependent beam control optimization of a traveling wave linear accelerator to improve VMAT delivery
Adriaan A van Appeldoorn1, Johannes G M Kok1, Jochem W H Wolthaus1
1Department of Radiotherapy, University Medical Center Utrecht, Utrecht, The Netherlands.
Introduction:
Increased modulation and dynamical delivery of external beam radiotherapy (EBRT), such as volumetric modulated arc therapy (VMAT) with dynamic gantry rotation, continuously variable dose rate (CVDR) and field shapes that change during the beam, place greater demands on the performance of linear accelerators (linac). In this study, the accuracy of the linac beam steering is improved by the application of a new method to determine the gantry-dependent lookup table.
Methods:
An improved method of lookup table creation based on service graphing information from the linac is investigated. This minimizes the impact of magnetic hysteresis due to the previous current in the steering magnets, which is dependent on the previous gantry angle. A software tool, programmed with MATLAB®, is used to calculate and export the new optimal lookup table (LUT).
Results:
This method is efficient requiring little clinical machine time or analysis time, and leads to an improved VMAT delivery with a reduction of about 60 percent in beam steering errors. If the surrounding magnetic field is changed, for example, ramping a nearby magnetic resonance imaging system (MRI), the beam steering LUT optimization can be quickly performed.
Conclusion:
This study shows an improved linac stability using improved lookup tables. Resulting in a lower number of interruptions, preventing down-time, and a lower risk of intrafraction motion due to longer treatment times.
More Related Videos
06:40Automated Delivery of Microfabricated Targets for Intense Laser Irradiation Experiments
Published on: January 28, 2021
09:01Gain-compensation Methodology for a Sinusoidal Scan of a Galvanometer Mirror in Proportional-Integral-Differential Control Using Pre-emphasis Techniques
Published on: April 4, 2017
Related Concept Videos
Beams with Unsymmetric Loadings
The first moment-area theorem determines the slope at any point on the beam. This theorem indicates that the change in slope between two points on a beam...
Beams with Symmetric Loadings
The M/EI...
Impact Loading on a Cantilever Beam
When an object is dropped onto the free end of a cantilever, its potential energy due to gravity is...
Shear on the Horizontal Face of a Beam Element
Unsymmetric Bending - Angle of Neutral Axis
When a bending moment is applied at an angle θ concerning the vertical axis of a symmetrical member, it can be resolved into components along the member's principal...
Deflection of a Beam
Singularity functions, described in an earlier lesson, are powerful mathematical tools that represent discontinuities within a function commonly encountered in structural loading...