Inter- and intra-unit reliability of the COSMED K5: Implications for multicentric and longitudinal testing
Kay Winkert1, Rupert Kamnig1, Johannes Kirsten1
1Division of Sports and Rehabilitation Medicine, Ulm University, Ulm, Baden-Württemberg, Germany.
Purpose:
To evaluate the intra-unit (RELINTRA) and inter-unit reliability (RELINTER) of two structurally identical units of the metabolic analyser K5 (COSMED, Rome, Italy) that allows to utilize either breath-by-breath (BBB) or dynamic mixing chamber (DMC) technology.
Methods:
Identical flow- and gas-signals were transmitted to both K5s that always operated simultaneously either in BBB- or DMC-mode. To assess RELINTRA and RELINTER, a metabolic simulator was applied to simulate four graded levels of respiration. RELINTRA and RELINTER were expressed as typical error (TE%) and Intraclass Correlation Coefficient (ICC). To assess also inter-unit differences via natural respiratory signals, 12 male athletes performed one incremental bike step test each in BBB- and DMC-mode. Inter-unit differences within biological testing were expressed as percentages.
Results:
In BBB, TE% of RELINTRA ranged 0.30-0.67 vs. RELINTER 0.16-1.39 and ICC ranged 0.57-1.00 vs. 0.09-1.00. In DMC, TE% of RELINTRA ranged 0.38-0.90 vs. RELINTER 0.03-0.86 and ICC ranged 0.22-1.00 vs. 0.52-1.00. Mean inter-unit differences ranged -2.30-2.20% (Cohen's ds (ds) 0.13-1.52) for BBB- and -0.55-0.61% (ds 0.00-0.65) for DMC-mode, respectively. Inter-unit differences for [Formula: see text] and RER were significant (p < 0.05) at each step.
Conclusion:
Two structurally identical K5-units demonstrated accurate RELINTRA with TE < 2.0% and similar RELINTER during metabolic simulation. During biological testing, inter-unit differences for [Formula: see text] and RER in BBB-mode were higher than 2% with partially large ES in BBB. Hence, the K5 should be allocated personally wherever possible. Otherwise, e.g. in multicenter studies, a decrease in total reliability needs to be considered especially when the BBB-mode is applied.
More Related Videos
08:40Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
09:18Author Spotlight: Assessing the Reliability of Doppler Ultrasound in Measuring Leg Blood Flow
Published on: December 15, 2023
Related Concept Videos
Reliability and Validity
Drug Product Performance: In Vitro–In Vivo Correlation
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...