Volume Elastic Modulus, Vascular Function, and Vascular Structure in Patients with Cardiovascular Risk Factors
Tatsuya Maruhashi1, Masato Kajikawa2, Shinji Kishimoto1
1Department of Cardiovascular Regeneration and Medicine, Research Institute for Radiation Biology and Medicine, Hiroshima University.
Aims:
Volume elastic modulus (VE), an index of arterial elasticity, and arterial diameter of the brachial artery can be automatically measured by a newly developed oscillometric device. We investigated the associations of VE with flow-mediated vasodilation (FMD), an index of endothelium-dependent vasodilation, nitroglycerine-induced vasodilation (NID), an index of endothelium-independent vasodilation, and intima-media thickness (IMT) of the brachial artery and association of oscillometrically measured brachial artery diameter with ultrasonographically measured brachial artery diameter in patients with cardiovascular risk factors.
Methods:
Oscillometric measurements of VE and brachial artery diameter and ultrasound measurements of brachial artery diameter, FMD, NID, and IMT of the brachial artery were performed in 50 patients with cardiovascular risk factors.
Results:
The mean values were 2.1±0.4 mmHg/% for VE, 0.31±0.05 mm for brachial IMT, 4.48±0.70 mm for oscillometric brachial artery diameter, and 4.30±0.55 mm for ultrasound brachial artery diameter. VE significantly correlated with brachial IMT (r=0.51, P<0.001), whereas there was no significant correlation of VE with FMD (r=-0.08, P=0.58) or NID (r=0.07, P=0.61). Multivariate analysis revealed that VE was significantly associated with brachial IMT (β=0.33, P=0.04). Oscillometric brachial artery diameter significantly correlated with ultrasound brachial artery diameter (r=0.79, P<0.001). The Bland-Altman plot showed good agreement between oscillometric brachial artery diameter and ultrasound brachial artery diameter (mean difference, -0.17 mm; limits of agreement, -1.03 mm to 0.69 mm).
Conclusions:
In patients with cardiovascular risk factors, VE may represent atherosclerotic structural alterations of the vascular wall but not vascular function. The accuracy of oscillometric measurement of brachial artery diameter is acceptable.
Related Concept Videos
Vascular Resistance
The primary determinants of vascular resistance are vessel diameter, blood viscosity, and vessel length. Among these, vessel diameter plays the most significant role due to the fourth power relationship described by...
Structure of Blood Vessels
Arteries and Arterioles
Vascular Spasm
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...
Venous Return
What is Venous Return?
Venous return refers to the rate at which blood flows back to the heart from the body's peripheral veins. It's an integral part of the circulatory system...