Accuracy of tidal volume delivery by five different models of large-animal ventilators
Objective:
To determine the accuracy of tidal volume (VT) delivery among 5 different models of large-animal ventilators when tested at various settings for VT delivery, peak inspiratory flow (PIF) rate, and fresh gas flow (FGF) rate.
Sample:
4 different models of pneumatically powered ventilators and 1 electrically powered piston-driven ventilator.
Procedures:
After a leak flow check, each ventilator was tested 10 times for each experimental setting combination of 5 levels of preset VT, 3 PIF rates, and 4 FGF rates. A thermal mass flow and volume meter was used as the gold-standard method to measure delivered VT. In addition, circuit systems of rubber versus polyvinyl chloride breathing hoses were evaluated with the piston-driven ventilator. Differences between preset and delivered VT (volume error [ΔVT]) were calculated as a percentage of preset VT, and ANOVA was used to compare results across devices. Pearson correlation coefficient analyses and the coefficient of determination (r2) were used to assess potential associations between the ΔVT and the preset VT, PIF rate, and FGF rate.
Results:
For each combination of experimental settings, ventilators had ΔVT values that ranged from 1.2% to 22.2%. Mean ± SD ΔVT was 4.8 ± 2.5% for the piston-driven ventilator, compared with 6.6 ± 3.2%, 10.6 ± 2.9%, 13.8 ± 2.97%, and 15.2 ± 2.6% for the 4 pneumatic ventilators. The ΔVT increased with higher PIF rates (r2 = 0.69), decreased with higher FGF rates (r2 = 0.62), and decreased with higher preset VT (r2 = 0.58).
Conclusions And Clinical Relevance:
Results indicated that the tested ventilators all had ΔVT but that the extent of each of ΔVT varied among ventilators. Close monitoring of delivered VT with external flow and volume meters is warranted, particularly when pneumatic ventilators are used or when very precise VT delivery is required.
More Related Videos
06:09Author Spotlight: Exploring Venous Waveforms in Porcine Models to Tackle Volume Overload in Medicine
Published on: January 12, 2024
06:22Surfactant Depletion Combined with Injurious Ventilation Results in a Reproducible Model of the Acute Respiratory Distress Syndrome ARDS
Published on: April 7, 2021
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Respiratory Volumes and Capacities