THE VALIDITY AND RELIABILITY OF THE HANDHELD OXIMETER TO DETERMINE LIMB OCCLUSION PRESSURE FOR BLOOD FLOW RESTRICTION
Anders Falk Brekke, Anders Nordahl Sørensen1, Christel Buhr1
1University College Absalon, Center of Nutrition and Rehabilitation, Department of Physiotherapy, Region Zealand, Denmark.
Background:
The limb occlusion pressure (LOP) is determined to calculate the relative LOP. The different levels of relative LOP (percentage of LOP) influence the treatment effect and perceived discomfort during low-load blood flow restriction (BFR) strength training. Thus, determining the LOP is of the utmost importance when using BFR in clinical practice.
Purpose:
The objective of this study was to investigate the concurrent validity and intra-rater (test-retest), intra-day reliability of an inexpensive, portable, easy-to-use handheld (HH) oximeter compared to a high-resolution Doppler ultrasound scanner in detecting LOP in the lower extremity.
Study Design:
Cross-sectional validity and reliability study.
Methods:
Two raters who were blinded from each other simultaneously assessed 50 healthy participants (mean age of 25.8 years). A 20 cm-wide thigh cuff with an attached sphygmomanometer was inflated until the raters independently registered the LOP with the HH oximeter and the Doppler ultrasound scanner. The test session was repeated once after a five-minute time interval.
Results:
The HH oximeter recorded a non-significantly higher LOP than the Doppler ultrasound scanner, with a mean difference of 6.3 mmHg in the test session (95% limits of agreement (LoA): -16.2 to 28.8, p = 0.13) and 5.4 mmHg in the retest session (95% LoA: -13.3 to 24.0, p = 0.10). The intra-rater reliability for both devices was moderate (ICC = 0.72-0.79). The measured LOP was significantly lower (p < 0.005) in the retest session than in the test session for both the HH oximeter (mean difference: -5.7 mmHg) and the Doppler ultrasound scanner (mean difference: -4.8 mmHg).
Conclusions:
The HH oximeter is a valid and reliable measuring device for determining the LOP in the lower extremity in healthy adults. The authors recommend performing at least two LOP measurements with a one-minute rest interval.
Level Of Evidence:
2, Validity and reliability study.
More Related Videos
04:44Author Spotlight: Enhancing Vascular Function and Physical Capacity in Cardiovascular Disease Through Novel Interventions and NIRS Technology
Published on: March 22, 2024
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessing Blood pressure in the Leg
Preparation:
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Pre-Procedural Guidelines for Assessing Blood Pressure
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: