Related Experiment Video
Updated: Aug 10, 2026

Development and Validation of an Ultrasensitive Single Molecule Array Digital Enzyme-linked Immunosorbent Assay for Human Interferon-α
Published on: June 14, 2018
Immunonephelometry of apolipoprotein B with a centrifugal analyzer
N Q Hanson1, T S LeGeault, E F Freier
1Department of Laboratory Medicine and Pathology, University of Minnesota Medical School, Minneapolis.
Abstract:
We developed an automated immunonephelometric assay for the measurement of apolipoprotein B (apo B) with a light-scattering microcentrifugal analyzer. Pretreating specimens with a dilute solution of Tween 20 or triglyceride lipase decreased the nephelometric response of apo B. Polyethylene glycol is included in the reaction mixture, and the reaction is complete within 4 min. The method is precise (CV = 6.5%, mean = 0.68 g/L) and the standard curve is linear to an apo B concentration of 2.8 g/L. Lipemia does not interfere with the method if grossly lipemic specimens are centrifuged to remove chylomicrons.

