Novel Ultrafast Spiral Head MR Angiography Compared to Standard MR and CT Angiography
Tobias Greve1,2, Nico Sollmann1,3, Andreas Hock4
1Department of Diagnostic and Interventional Neuroradiology, Klinikum rechts der Isar, Technische Universität München, Munich, Germany.
Background And Purpose:
Intracranial vessel imaging by time-of-flight MR angiography (TOF-MRA) is one of the most frequently performed investigations in clinical neuroradiology. Particularly in the acute setting, fast imaging is needed for diagnostics, with a sequence ideally depicting even small vessels. The purpose of this study was to compare image and diagnostic quality of a novel ultrashort TOF-MRA sequence accelerated by spiral imaging (TOF-Spiral-short) to a standard TOF-MRA sequence accelerated by compressed sensing (TOF-CS) and to CT angiography (CTA).
Methods:
Forty-one patients (36.6% showing vessel pathologies) who had undergone TOF-CS (acquisition duration: 4 minutes 8 seconds), TOF-Spiral-short (acquisition duration: 51 seconds; spiral imaging [accelerating factor 1.3], decreased field of view [accelerating factor 1.2], and increased voxel size [accelerating factor 3.3]), and CTA were retrospectively evaluated. Assessment of image quality, diagnostic confidence, and quantification of stenosis or aneurysm diameter were performed by two readers.
Results:
Image quality at the skull base was slightly reduced with TOF-Spiral-short compared to CTA and TOF-CS (P < .05). Delineation of small intracranial vessels was improved by TOF-Spiral-short compared to CTA (P < .0001). In TOF-Spiral-short, diagnostic confidence was not reduced compared to TOF-CS in patients with vessel pathologies. We observed no significant difference in quantitative pathology assessment between TOF-Spiral-short and the other two modalities. TOF-Spiral-short enabled the correct identification of all vessel pathologies.
Conclusions:
Accelerating TOF-MRA of brain-feeding arteries by a novel ultrashort spiral imaging sequence shows adequate image quality and sufficient diagnostic performance. Thus, TOF-Spiral-short holds potential for fast and reliable diagnostics of vessel pathologies, particularly in the acute setting.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...