Automated Noninvasive Central Blood Pressure Measurements by Oscillometric Radial Pulse Wave Analysis: Results of the
Frederik Trinkmann1, Urs Benck2, Julian Halder1
11st Department of Medicine (Cardiology, Angiology, Pulmonary and Intensive Care), University Medical Center Mannheim, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany.
Background:
Central blood pressure becomes increasingly accepted as an important diagnostic and therapeutic parameter. Accuracy of widespread applanation tonometry can be affected by calibration and operator training. To overcome this, we aimed to evaluate novel VascAssist 2 using automated oscillometric radial pulse wave analysis and a refined multi-compartment model of the arterial tree.
Methods:
Two hundred and twenty-five patients were prospectively enrolled. Invasive aortic root measurements served as reference in MEASURE-cBP 1 (n = 106) whereas applanation tonometry (SphygmoCor) was used in MEASURE-cBP 2 (n = 119).
Results:
In MEASURE-cBP 1, we found a mean overestimation for systolic values of 4 ± 12 mmHg (3 ± 10%) and 6 ± 10 mmHg (9 ± 14%) for diastolic values. Diabetes mellitus and low blood pressure were associated with larger variation. In MEASURE-cBP 2, mean overestimation of systolic values was 4 ± 4 mmHg (4 ± 4%) and 1 ± 4 mmHg (1 ± 7%) of diastolic values. Arrhythmia was significantly more frequent in invalid measurements (61 vs. 18%, P < 0.0001) which were most often due to a low quality index of SphygmoCor.
Conclusions:
Central blood pressure estimates using VascAssist 2 can be considered at least as accurate as available techniques, even including diabetic patients. In direct comparison, automated measurement considerably facilitates application not requiring operator training and can be reliably applied even in patients with arrhythmias.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)