Comparison of PR, QRS, and QT interval measurements by seven ECG interpretation programs
J De Bie1, I Diemberger2, J W Mason3
1Mortara Instrument Europe s.r.l., Bologna, Italy.
Digital electrocardiogram (ECG) programs show small systematic biases and potential for large errors in interval measurements. Users should exercise caution with automated ECG analysis, especially in complex cases.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Instrumentation
Background:
- Automated electrocardiogram (ECG) interval measurements are presumed more accurate than manual methods.
- However, performance variations among different ECG analysis programs remain under-evaluated.
Purpose of the Study:
- To compare the accuracy and consistency of PR, QRS, and QT interval measurements across multiple commercial electrocardiograph programs.
- To identify systematic biases and the extent of clinically significant errors generated by these automated ECG analysis tools.
Main Methods:
- Digital ECGs were replayed through seven commercial electrocardiographs to generate analog ECGs.
- Measurements for PR interval, QRS duration, and QT interval were extracted and compared against a median of program outputs and population mean values.
Main Results:
- Small, statistically significant systematic biases were observed between different ECG measurement programs.
- While most pairwise comparisons showed similar accuracy for routine ECGs, substantial differences emerged in the frequency and magnitude of clinically significant errors, particularly in challenging cases.
- Variations in error magnitudes (e.g., at the 98th percentile) could differ by a factor of two, with inconsistencies in overestimation and underestimation patterns.
Conclusions:
- Users of digital ECGs must acknowledge the presence of systematic differences between analysis programs.
- Awareness of potential large, clinically relevant errors in difficult ECG tracings is crucial for accurate interpretation.
Related Concept Videos
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
ECG Interpretation of Arrhythmias I: Sinus Arrhythmias
Types of Arrhythmias
Sinus Node Arrhythmias
Sinus Bradycardia: Originating from the sinoatrial (SA) node, sinus bradycardia involves slower impulses, resulting in a heart rate of less than 60 beats per minute (bpm). Causes include sleep, vagal stimulation, beta-blockers, hypothyroidism,...