Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
Esophageal deviation with vacuum suction and mechanical deflection during ablation of atrial fibrillation: First in
Luis Aguinaga1, Alejandro Palazzo1, Alejandro Bravo1
1Centro Integral de Arritmias Tucumán, San Miguel de Tucumán, Tucumán, Argentina.
Abstract:
Deviation of the esophagus during atrial fibrillation (AF) ablation can reduce esophageal injury. This study reports upon a novel esophageal retractor that utilizes vacuum suction and mechanical deflection to deviate the esophagus. The device was used in seven patients undergoing cryoballoon AF ablation. The esophagus was deviated 31.9 ± 4.4 mm to the right and 28.2 ± 5.9 mm to the left. Endoscopy at 4.4 ± 1.5 days postablation showed no esophageal injury. This study demonstrates the safe and effective deviation of the esophagus without a trailing edge with an esophageal retractor utilizing vacuum suction and mechanical deflection.