Comparison of image quality and radiation dose using different pre-ASiR-V and post-ASiR-V levels in coronary computed
Yongxia Zhao1, Dongxue Li1, Zhichao Liu1
1Department of Radiology, The Affiliated Hospital of Hebei University, Baoding, China.
Objective:
To determine the optimal pre-adaptive and post-adaptive level statistical iterative reconstruction V (ASiR-V) for improving image quality and reducing radiation dose in coronary computed tomography angiography (CCTA).
Methods:
The study was divided into two parts. In part I, 150 patients for CCTA were prospectively enrolled and randomly divided into 5 groups (A, B, C, D, and E) with progressive scanning from 40% to 80% pre-ASiR-V with 10% intervals and reconstructing with 70% post-ASiR-V. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. Subjective image quality was assessed using a 5-point scale. The CT dose index volume (CTDIvol) and dose-length product (DLP) of each patient were recorded and the effective radiation dose (ED) was calculated after statistical analysis by optimizing for the best pre-ASiR-V value with the lowest radiation dose while maintaining overall image quality. In part II, the images were reconstructed with the recommended optimal pre-ASiR-V values in part I (D group) and 40%-90% of post-ASiR-V. The reconstruction group (D group) was divided into 6 subgroups (interval 10%, D0:40% post-ASiR-V, D1:50% post - ASiR-V, D2:60% post-ASiR-V, D3:70% post-ASiR-V, D4:80% post-ASiR-V, and D5:90% post-ASiR-V).The SNR and CNR of D0-D5 subgroups were calculated and analyzed using one-way analysis of variance, and the consistency of the subjective scores used the k test.
Results:
There was no significant difference in the SNRs, CNRs, and image quality scores among A, B, C, and D groups (P > 0.05). The SNR, CNR, and image quality scores of the E group were lower than those of the A, B, C, and D groups (P < 0.05). The mean EDs in the B, C, and D groups were reduced by 7.01%, 13.37%, and 18.87%, respectively, when compared with that of the A group. The SNR and CNR of the D4-D5 subgroups were higher than the D0-D3 subgroups, and the image quality scores of the D4 subgroups were higher than the other subgroups (P < 0.05).
Conclusion:
The wide-detector combined with 70% pre-ASiR-V and 80% post-ASiR-V significantly reduces the radiation dose of CCTA while maintaining overall image quality as compared with the manufacture's recommendation of 40% pre-ASiR-V.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...