A novel method for continuous blood pressure estimation based on a single-channel photoplethysmogram signal
Qihan Hu1,2, Xintao Deng3,2, Aiguo Wang3
1Department of Electronic Engineering, Fudan University, Shanghai 200433, People's Republic of China.
Objective:
Currently, continuous blood pressure (BP) measurements are mostly based on multi-sensor combinations and datasets with limited BP ranges. Besides, most BP-related features derive from the photoplethysmogram (PPG) signal. The mechanism of PPG signal formation is not considered. We aimed to design a noninvasive and continuous method for estimation of BP using a single PPG sensor, which takes the mechanism of PPG signal formation into account.
Approach:
We prepared a dataset containing PPG signals for 294 patients from three public databases for constructing the BP estimation model. The features used in the model consisted of two types: novel features based on a multi-Gaussian model and existing features. The multi-Gaussian model fitted the different components (i.e. the main wave, the dicrotic wave and the tidal wave) of the PPG signal. Ensemble machine learning algorithms were applied to estimate systolic blood pressure (SBP) and diastolic blood pressure (DBP). When partitioning the dataset, there was an overlap between the training set and the testing set.
Main Results:
Datasets with a wide-range of SBP and DBP values (SBP ranging from 74 to 229 mmHg and DBP ranging from 26 to 141 mmHg) were used to evaluate our method. The mean and standard deviation of error for SBP and DBP estimations were -0.21 ± 5.21 mmHg and -0.19 ± 3.37 mmHg, respectively. The model performance fully met the Association for the Advancement of Medical Instrumentation standard and was grade 'A' on the British Hypertension Society standard.
Significance:
The multi-Gaussian model could be used to estimate BP, and our method was able to track a wide range of BP accurately. In addition our method is based on a single PPG sensor, making it very convenient.
More Related Videos
08:22Author Spotlight: Advancing the Study of Brain-Heart Interplay with a Comprehensive EEGLAB Plugin for Multimodal Signal Analysis
Published on: April 26, 2024
11:26Assessing Cerebral Autoregulation via Oscillatory Lower Body Negative Pressure and Projection Pursuit Regression
Published on: December 10, 2014
Related Concept Videos
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.