Non-invasive measurement of pulse pressure variation using a finger-cuff method in obese patients having laparoscopic
Moritz Flick1, Roman Schumann2, Phillip Hoppe1
1Department of Anesthesiology, Center of Anesthesiology and Intensive Care Medicine, University Medical Center Hamburg-Eppendorf, Martinistrasse 52, 20246, Hamburg, Germany.
Abstract:
Pulse pressure variation (PPV) is a dynamic cardiac preload variable used to predict fluid responsiveness. PPV can be measured non-invasively using innovative finger-cuff systems allowing for continuous arterial pressure waveform recording, e.g., the Nexfin system [BMEYE B.V., Amsterdam, The Netherlands; now Clearsight (Edwards Lifesciences, Irvine, CA, USA)] (PPVFinger). However, the agreement between PPVFinger and PPV derived from an arterial catheter (PPVART) in obese patients having laparoscopic bariatric surgery is unknown. We compared PPVFinger and PPVART at 6 time points in 60 obese patients having laparoscopic bariatric surgery in a secondary analysis of a prospective method comparison study. We used Bland-Altman analysis to assess absolute agreement between PPVFinger and PPVART. The predictive agreement for fluid responsiveness between PPVFinger and PPVART was evaluated across three PPV categories (PPV < 9%, PPV 9-13%, PPV > 13%) as concordance rate of paired measurements and Cohen's kappa. The overall mean of the differences between PPVFinger and PPVART was 0.5 ± 4.6% (95%-LoA - 8.6 to 9.6%) and the overall predictive agreement was 72.4% with a Cohen's kappa of 0.53. The mean of the differences was - 0.7 ± 3.8% (95%-LoA - 8.1 to 6.7%) without pneumoperitoneum in horizontal position and 1.1 ± 4.8% (95%-LoA - 8.4 to 10.5%) during pneumoperitoneum in reverse-Trendelenburg position. The absolute agreement and predictive agreement between PPVFinger and PPVART are moderate in obese patients having laparoscopic bariatric surgery.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure in the Leg
Preparation: