Evaluating Tidal Volume Delivered by a Self-Inflating Bag Depending on the Compression Method
Woo Jin Cho1, Hyun Jung Kim2, Tae Ki Yang3
1Department of Anesthesiology and Pain Medicine, Jeju National University Hospital, Jeju, Republic of Korea.
Background:
A self-inflating bag (SIB) is an essential device for pulmonary resuscitation in emergencies. However, it is difficult for a rescuer to accurately predict a tidal volume (TV) delivered by a SIB to a patient.
Objective:
The objective of this study was to evaluate TV values delivered by a SIB with various compression methods.
Methods:
A total of 33 participants were enrolled in this study. Anterior, middle, and posterior points were marked on the surface of the SIB to guide participants to place their fingers. This was connected to a gas-flow analyzer and a test lung to measure the TV. Participants were asked to squeeze the SIB with one-handed and two-handed compression at anterior, middle, and posterior points.
Results:
The TV was the largest when the bag was squeezed with two-handed compression at the middle points (578 ± 21 mL). It was the smallest when the bag was squeezed with one-handed compression at the anterior point (311 ± 15 mL). Except that the TV with one-handed compression at the middle point and the TV with two-handed compression at the posterior points were similar, all other TV values were significantly different from each other (all p < 0.001).
Conclusion:
The TV delivered by a SIB could differ greatly depending on the compression points and whether one or two hands are used. Therefore, a rescuer should choose an appropriate compression method considering patient's physical characteristics and medical conditions.
More Related Videos
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
Related Concept Videos
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...