Published on: September 18, 2012
Validation of red blood cell flux and velocity estimations based on optical coherence tomography intensity
Paul J Marchand1,2, Xuecong Lu3,4, Cong Zhang3,4
1Department of Electrical Engineering, Polytechnique Montreal, Montreal, Canada. pjsmarchand@gmail.com.
Abstract:
We present a validation of red blood cell flux and speed measurements based on the passage of erythrocytes through the OCT's focal volume. We compare the performance of the so-called RBC-passage OCT technique to co-localized and simultaneously acquired two-photon excitation fluorescence microscopy (TPEF) measurements. Using concurrent multi-modal imaging, we show that fluctuations in the OCT signal display highly similar features to TPEF time traces. Furthermore, we demonstrate an overall difference in RBC flux and speed of 2.5 ± 3.27 RBC/s and 0.12 ± 0.67 mm/s (mean ± S.D.), compared to TPEF. The analysis also revealed that the OCT RBC flux estimation is most accurate between 20 RBC/s to 60 RBC/s, and is severely underestimated at fluxes beyond 80 RBC/s. Lastly, our analysis shows that the RBC speed estimations increase in accuracy as the speed decreases, reaching a difference of 0.16 ± 0.25 mm/s within the 0-0.5 mm/s speed range.